Product Information:

The BINDER incubator of the BF series with mechanical convection is suitable for all gentle incubation applications, particularly under a full load and large batch throughputs. This incubator is extremely homogeneous and has quick recovery times

Performance characteristics
  • Electronically controlled APT.line™ preheating chamber assuring temperature accuracy and reproducible results 
  • Temperature range from 5 °C above room temperature to 100 °C 
  • Digital temperature setting with an accuracy of a tenth of a degree 
  • MS controller with several timer functions 
  • Controller time functions: 
  • 0 to 99 h 59 min delayed ON 
  • 0 to 99 h 59 min delayed OFF 
  • 0 to 99 h 59 min temperature-dependent 
  • delayed OFF 
  • Adjustable fan speed 
  • Adjustable ventilation by means of front ventilation flap slide and rear exhaust Ø 50 mm 
  • Inner glass door 
  • Independent temperature safety device class 3.1 (DIN 12880) with optical alarm 
  • RS 422 interface for use with APT-COM™ DataControlSystem communication software or switch over to printer output with RS 232 / RS 422 interface converter 
  • Adjustable intervals for printer 
  • Units up to 115 liters are stackable 
  • 2 chrome-plated racks included 
  • BINDER test confirmation 
Options
  • Access ports with silicone plug, 10, 30, 50, 100 mm 
  • Switchable audible alarm for over temperature. Adjustable limit on the independent temperature safety device 
  • Switchable waterproof interior socket 230 V AC (max. 500 W), protection class IP 65, with corresponding plug. Max. permissible operating temperature 50 °C 
  • Additional PT 100 temperature sensor, flexibly installed with external connection, includes LEMO connector (3-pin) 
  • Temperature measurement acc. to DIN 12880 (27 measuring points) at 37 °C or at specified temperature with measuring protocol and certificate 
  • Calibration certificate, measurement in center of chamber at 37 °C or at specified testing temperature 
  • Extension to calibration certificate. Each additional measurement at additional measuring point or testing temperature 
  • Anti-slip rubber pads for safe stacking (1 set of 4 pieces) 
  • Data Logger Kit T 220: For continuous temperature recording of -90 °C to 220 °C. Kit includes 1 data logger, PT 100 sensor with 2 m Teflon extension cable and 1 fixture for mounting to the BINDER unit 
  • Data Logger Software: Configuration and evaluation software for all BINDER Data Logger Kits, incl. data cable 
  • Rack, chrome-plated 
  • Rack, stainless steel 
  • Perforated shelf, stainless steel 
  • Door lock 
Please contact sales office for quotation

View FileBinder Product Catalogue    Size: (10.69 MB)
View FileBinder Company Brochure    Size: (3.91 MB)
View FileKaisertech Product Guide v.2.o    Size: (5.76 MB)

General Specifications

Exterior dimensions 
  • Width (mm) 1235 
  • Height (incl. feet) (mm) 1025 
  • Depth (mm) 765 
  • Plus door handle and connection (mm) 90 
  • Wall clearance, rear (mm) 100 
  • Wall clearance, side (mm) 160 
  • Exhaust duct (outer Ø mm) 52 
  • Steam space volume (l) 457 
  • Number of doors (ea.) 2 
Interior dimensions 
  • Width (mm) 1000 
  • Height (mm) 800 
  • Depth (mm) 500 
  • Interior volume (l) 400 
  • Racks (number standard/max.) 2 / 10 
  • Load per rack (kg) 35 
  • Permitted total load (kg) 90 
  • Weight (empty) (kg) 145 
Temperature data 
  • Temperature range approx. 5 °C above ambient temperature to (°C) 100 
  • Temperature variation 
    • at 37 °C (± K) 0,4 
    • at 50 °C (± K) 0,9 
  • Temperature fluctuation 
    • at 37 °C (± K) 0,2 
    • at 50 °C (± K) 0,2 
  • Warm-up time 1) 
    • to 37 °C (min.) 18 
    • to 50 °C (min.) 26 
  • Recovery time after doors were open for 30 sec. 1) 
    • at 37 °C (min.) 2 
    • at 50 °C (min.) 4 
  • Ventilation at 70 °C (x/h) 17 
Electrical data 
  • IP protection class acc. to EN 60529 IP 20 
  • Voltage (± 10%) 50 / 60 Hz (V) 230 
  • Nominal power (kW) 0,85 
  • Energy consumption at 37 °C (W) 56 
1) To 98% of the set value 

All technical data are specified for units with standard equipment at an ambient temperature of 25 °C and a line voltage fluctuation of ±10%. The temperature data is determined in accordance to factory standard following DIN 12880, respecting the recommended wall clearances of 10% of the height, width and depth of the inner chamber. All figures are typical average values for series devices. We reserve the right to alter technical specifications at any time