Product Information:

The BINDER FDL 115 with silicone and dust-free inner chamber, as well as symmetric airflow dries specimens containing solvents and thus this safety drying oven meets all EN 1539 requirements.

Please contact sales office for quotation

Performance Charateristics
  • Electronically controlled APT.line™ preheating chamber assuring temperature accuracy and reproducible results 
  • All safety features met according to EN 1539 
  • Temperature range from 5 °C above ambient temperature to 300 °C 
  • MP controller with 2 programs with 10 sections each, or alternatively one program with 20 sections 
  • The time of an individual program step can be set to max. 99.59 hours or 999.59 hours. This adjustment applies to all program sections. 
  • Integrated weekly program timer with real-time function 
  • Adjustable ramp function via program editor 
  • Digital temperature setting with an accuracy of one degree 
  • Elapsed time indicator 
  • Door gasket FKM (for max. temperature of 200 °C / 392 °F) 
  • Replaceable fresh-air filter cartridge. Class F6 (EU6 fine-particle filter for particle sizes 1 µm to 10 µm) 
  • Independent adjustable temperature safety device class 2 (DIN 12880), with visual temperature alarm 
  • Fresh-air monitoring with visual and audible alarm and automatic deactivation of heating 
  • Rear exhaust connector Ø 100 mm (4 inch) 
  • RS 422 interface for use with APT-COM™ DataControlSystem communication software 
  • Units up to 115 liters are stackable 
  • 2 chrome-plated racks included 
  • BINDER test confirmation 
Options 
  • Access ports with silicone plug, 10, 30 mm 
  • Rack,chrome-plated or stainless steel 
  • Perforated shelf, stainless steel 
  • Door lock 
  • Door gasket made of of high temperature resistant silicone for operating temperatures > 200 °C 
  • Specimen temperature display with clip temperature sensor and 4 - 20 mA analog output (e.g. chart recorder connection), accessible via 6-pin DIN socket 
  • Calibration certificate, measurement in center of chamber at 150 °C or at specified testing temperature 
  • Extension to calibration certificate. Each additional measurement at additional measuring point or testing temperature 
  • Temperature measurement acc. to DIN 12880 (27 measuring points) at 150 °C or at specified temperature with measuring protocol and certificate 
  • Replacement air filter (class F6/EU6 – for particle sizes between 1 µm and 10 µm), 100x520x22 mm 3.94x20.47x0.87 inch), with aluminum frame
View FileBinder Product Catalogue    Size: (10.69 MB)
View FileBinder Company Brochure    Size: (3.91 MB)
View FileKaisertech Product Guide v.2.o    Size: (5.76 MB)

General Specifications

Exterior dimensions 
  • Width (mm) 835 
  • Height (incl. feet) (mm) 800 
  • Depth (plus door handle 50 mm) (mm) 685 
  • Wall clearance, rear (mm) 100 
  • Wall clearance, side (mm) 100 
  • Exhaust duct (outer Ø mm) 100 
  • Steam space volume (l) 156 
Interior dimensions 
  • Width (mm) 600 
  • Height (mm) 435 
  • Depth (mm) 435 
  • Interior volume (l) 115 
  • Racks (number standard/max.) 2 / 5 
  • Load per rack (kg) 20 
  • Permitted total load (kg) 50 
  • Weight (empty) (kg) 90 
Temperature data 
  • Temperature range approx. 5 °C above ambient temperature to (°C) 300 
  • Temperature variation 
    • at 70 °C (± K) 1,5 
    • at 150 °C (± K) 2,5 
    • at 300 °C (± K) 4 
  • Temperature fluctuation (± K) 0,3 
  • Warm-up time 1) 
    • to 70 °C (min.) 7 
    • to 150 °C (min.) 17 
    • to 300 °C (min.) 44 
  • Recovery time after doors were open for 30 sec. 1) 
    • at 70 °C (min.) 1 
    • at 150 °C (min.) 3 
    • at 300 °C (min.) 6 
Ventilation data 
  • Ventilation (approx. x/min.) 3 
  • Air circulation (approx. x/min.) 40 
  • Exhaust air volume flow (approx. l/min. m³/h) 400 (24,0) 
  • Air flow velocity (m/sec) 0,8 - 1,2 
  • Highest permitted solvent quantity (at T-180 °C, M-100g/mol, U-40g/m3, K=0.5) (g) 6,65 
Electrical data 
  • IP protection class acc. to EN 60529 IP 33 
  • Voltage (± 10%) 50 / 60 Hz (V) 230 
  • Nominal power (kW) 2,9 
  • Energy consumption 
  • at 150 °C (W) 1098 
1) T = drying temperature
M = molecular mass
U = lower explosion limit
K = solvent vapor concentration as percentage of lower explosion limit

2) To 98% of the set value

All technical data are specified for units with standard equipment at an ambient temperature of 25 °C and a line voltage fluctuation of ±10%. The temperature data is determined in accordance to factory standard following DIN 12880, respecting the recommended wall clearances of 10% of the height, width and depth of the inner chamber. All figures are typical average values for series devices. We reserve the right to alter technical specifications at any time