Product Information:

FD series units are primarily used in applications needing rapid drying and sterilization. Totally homogenous temperature distribution, rapid dynamic response, and a special air turbine which was developed by us and has 20 % higher output, have made the FD series a genuine time-saving device.

Features
  • Electronically controlled APT.line™ preheating chamber assuring temperature accuracy and reproducible results 
  • Temperature range 5 °C (32 ºF) above ambient temperature up to 300 °C (572 °F) 
  • DS controller with integrated timer 0 to 99 h 
  • Digital temperature setting with an accuracy of one degree 
  • Independent adjustable temperature safety device class 2 (DIN 12880), with visual temperature alarm 
  • Adjustable front ventilation flap slide and rear exhaust ø 50 mm (1.97 inch) 
  • Units up to 115 liters are stackable 
  • 2 chrome-plated racks included 
  • BINDER test certificate 
Options
  • Access port with silicone plugs, 10 mm (0.39 inch), 30 mm (1.18 inch), 50 mm (1.97 inch), 100 mm (3.94 inch) 
  • Independent adjustable temperature safety device, Class 3.1 (DIN 12880) 
  • Analog temperature output, 4 - 20 mA, with 6 - pin DIN socket (output not adjustable) 
  • Over temperature alarm, acoustic, can be switched off. Temperature limit can be set at the independent, adjustable temperature safety device class 2 
  • Anti-slip rubber pads for safe stacking (4 pieces) 
  • Temperature measurement acc. to DIN 12880 (27 measuring points) at 150 °C (302 °F) or at specified temperature with measuring protocol and certificate 
  • Factory calibration certificate. Measurement in center of chamber at 150 °C (302 °F) or at specified testing temperature 
  • Extension to factory calibration certificate. Each additional measurement at an additional measuring point or temperature 
  • Rack, chrome - plated or stainless steel 
  • Shelf, perforated, stainless steel 
  • Lockable door 
  • FKM door gasket 
  • Door with window and interior lighting, 15 W (230 x 230 mm / 9.06 x 9.06 inch) 
View FileBinder Product Catalogue    Size: (10.69 MB)
View FileBinder Company Brochure    Size: (3.91 MB)
View FileKaisertech Product Guide v.2.o    Size: (5.76 MB)

General Specifications

Exterior dimensions 
  • Width (mm/inch) 634 / 25.0 
  • Height (inclusive feet/castors) (mm/inch) 617 / 24.3 
  • Depth (mm/inch) 575 / 22.6 
  • Plus door handle, I-panel and exhaust duct (mm/inch) 105 / 4.1 
  • Wall clearance rear (mm/inch) 100 / 3.9 
  • Wall clearance side (mm/inch) 160 / 6.3 
  • Exhaust duct outer- Ø (mm/inch) 52 / 2.1 
  • Steam space volume (l/cu.ft.) 77 / 2.7 
  • Number of doors 1 
Interior dimensions 
  • Width (mm/inch) 400 / 15.8 
  • Height (mm/inch) 400 / 15.8 
  • Depth (mm/inch) 330 / 13 
  • Interior volume (l/cu.ft.) 53 / 1.9 
  • Racks, chrome-plated (number standard/max.) 2 / 5 
  • Load per rack (kg/lbs.) 15 / 33 
  • Permitted total load (kg/lbs.) 40 / 88 
  • Weight of the unit (empty) (kg/lbs.) 44 / 97 
Temperature data 
  • Temperature range, 5°C (41°F) above ambient up to (°C / °F) 300 / 572 
  • Temperature variation 1) 
    • at 70 °C (± °C) 0,8 
    • at 150 °C (± °C) 2 
    • at 300 °C (± °C) 3.7 
  • Temperature fluctuation (± °C) 0.3 
  • Heating up time 1), 2) 
    • to 70 °C (Min.) 7 
    • to 150 °C (Min.) 22 
    • auf 300 °C (Min.) 45 
  • Recov. time after door was opened for 30 sec. 
    • at 70 °C (Min.) 2 
    • at 150 °C (Min.) 4 
    • at 300 °C (Min.) 9 
Air change data 
  • at 70 °C (x/h) 59 
  • at 150 °C (x/h) 64 
  • at 300 °C (x/h) 53 
Electrical data 
  • Housing protection acc. to EN 60529 IP 20 
  • Nominal voltage (±10 %) 50 / 60 Hz (V) 230 
  • Nominal power (kW) 1.2 
Energy consumption 
  • at 70 °C (W) 172 
  • at 150 °C (W) 429 
  • at 300 °C (W) 951 
1) value without window 
2) up to 98 % of the set value 
3) The air change depends on the inner chamber- and ambient temperature and is subject to significant individual variance. The indicated air change rate represents average values for standard equipment. Individual measurement of air change rate in acc. To ASTM D 5374 are optionally available.

All technical specification are specified for units with standard equipment at an ambient temperature of 25 °C (77 °F) and a voltage fluctuation of ±10 %. The temperature data are determinated in accordance to factory standard following DIN 12880, part 2 respecting the recommended wall clearances of 10 % of the height, width and depth of the inner chamber. All indications are average values, typical for units produced in series. We reserve the right to alter technical specifications at all times.

Write Your Own Review

You're reviewing: Binder - FD53 Drying Oven with forced convection

How do you rate this product? *

  1 star 2 stars 3 stars 4 stars 5 stars
Price
Value
Quality