Binder - M 400 Material Test Chamber With Individual Programming 9010-0204
Product Information
With a maximum temperature of 300 °C and comprehensive programming options, the M series from BINDER is ideally suited for material and aging testing. The high-performance fan ensures fast heating
Performance characteristics
Exterior dimensions
All technical data are specified for units with standard equipment at an ambient temperature of 25 °C and a line voltage fluctuation of ±10%. The temperature data is determined in accordance to factory standard following DIN 12880, respecting the recommended wall clearances of 10% of the height, width and depth of the inner chamber. All data was determined at 100% of fan speed. All figures are typical average values for series devices. We reserve the right to alter technical specifications at any time
With a maximum temperature of 300 °C and comprehensive programming options, the M series from BINDER is ideally suited for material and aging testing. The high-performance fan ensures fast heating
Performance characteristics
- Electronically controlled APT.line™ preheating chamber assuring temperature accuracy and reproducible results
- Temperature range from 5 °C above ambient temperature to 300 °C
- MCS controller with 25 storable programs of 100 sections each for a maximum of 500 program segments
- User-friendly LCD screen
- Easy-to-read menu guide
- Integrated electronic chart recorder
- Variety of options for the graphic display of process parameters
- Real-time clock
- Adjustable ramp function via program editor
- Program-controlled ventilation flap
- High air-exchange rate through high-performance fan
- Adjustable fan speed
- Exhaust duct Ø 50 mm
- Temperature safety device class 2 (DIN 12880) with visual alarm
- Printer and communication interface RS 422 for use with optional GMP/GLP and FDA guideline 21 CFR Part 11 compliant APT-COM™ DataControlSystem software
- Units up to 115 liters are stackable
- 2 chrome-plated racks included
- BINDER test confirmation
- Access ports with silicone plug, 10, 30, 50, 100 mm
- Securing elements for additional fastening of racks (1 set of 4 pieces)
- Anti-slip rubber pads for safe stacking (1 set of 4 pieces)
- Keyboard lock
- HEPA fresh-air filter, Class H 14 (according to EN 1822, min. 99.995% for 0.1-0.3 µm particles)
- Measurement of air change rate according to ASTM D5374, with definition and protocol according to ambient temperature (measurement in factory)
- Analog temperature output 4 - 20 mA with 6-pin DIN socket (output not adjustable)
- Additional PT 100 temperature sensor, flexibly installed with external connection, includes LEMO connector (3-pin)
- Additional measuring channel for digital display of specimen temperature with flexible PT 100 temperature sensor, measured data recorded via RS 422 interface
- Temperature measurement acc. to DIN 12880 (27 measuring points) at 150 °C or at specified temperature with measuring protocol and certificate
- Factory calibration certificate. Measurement in center of chamber at 150 °C (302 °F) or at specified testing temperature
- Extension to factory calibration certificate. Each additional measurement at additional measuring point or temperature
- Data Logger Kit T 350: For continuous temperature recording of 0 °C to 350 °C. Kit includes 1 data logger, PT 100 sensor with 2 m Teflon extension cable and 1 fixture for mounting to the BINDER unit
- Data Logger Software: Configuration and evaluation software for all BINDER Data Logger Kits, incl. data cable
- Rack, chrome-plated
- Rack, stainless steel
- Reinforced rack, stainless steel, with 1 set of fasteners (4 pieces), max. load 70 kg
- Shelf, perforated, stainless steel
- Locking door handle with key
- Door gasket, FKM (Viton)
- 2 doors each with window 470 x 290 mm and interior lighting (30 W)
- Reinforced inner chamber, including 2 reinforced racks (maximum total load 250 kg, max. load per rack 70 kg)
Exterior dimensions
- Width (mm) 1235
- Height (incl. feet) (mm) 1185
- Depth (mm) 765
- Plus door handle and connection (mm) 150
- Wall clearance, rear (mm) 100
- Wall clearance, side (mm) 160
- Exhaust duct (outer Ø mm) 52
- Steam space volume (l) 498
- Number of doors (ea.) 2
- Width (mm) 1000
- Height (mm) 800
- Depth (mm) 510
- Interior volume (l) 400
- Racks (number standard/max.) 2 / 10
- Load per rack (kg) 35
- Permitted total load (kg) 90
- Weight (empty) (kg) 173
- Temperature range approx. 5 °C above ambient temperature to (°C) 300
- Temperature variation
- at 70 °C (± K) 0,7
- at 150 °C (± K) 1,5
- at 300 °C (± K) 5
- Temperature fluctuation (± K) 0,3
- Warm-up time 1)
- to 70 °C (min.) 6
- to 150 °C (min.) 18
- to 250 °C (min.) 44
- Recovery time after doors were open for 30 sec. 1)
- at 70 °C (min.) 1
- at 150 °C (min.) 3
- at 300 °C (min.) 5
- at 70 °C (x/h) 51
- at 150 °C (x/h) 54
- at 300 °C (x/h) 48
- IP protection class acc. to EN 60529 IP 20
- Voltage (± 10%) 50/60 Hz (V) 400/3N~
- Nominal power (kW) 3,4
- at 70 °C (W) 520
- at 150 °C (W) 1200
- at 300 °C (W) 2340
All technical data are specified for units with standard equipment at an ambient temperature of 25 °C and a line voltage fluctuation of ±10%. The temperature data is determined in accordance to factory standard following DIN 12880, respecting the recommended wall clearances of 10% of the height, width and depth of the inner chamber. All data was determined at 100% of fan speed. All figures are typical average values for series devices. We reserve the right to alter technical specifications at any time