BINDER - KMF Series | Expanded Temperature/Humidity Range | Stress Testing
The BINDER KMF Series guarantees constant test conditions throughout the testing area of the oven. A large advantage of this series is its interior and exterior design.
The KMF ovens' have low space requirements within the humidity test chambers, alongside flexibility in terms of water supply. The wide temperature and humidity range make this humidity test chamber ideally suited for stress testing.
Finally, this range comes with a wide selection of accessories that allows the customer to adapt to their specific needs.
Key Features:
Stainless Steel Interior - The inner chamber and all fixtures are stainless steel. This means there is no risk of corrosion and cleaning after use is easy.
Homogeneous Temperature Distribution - BINDER's APT.line preheating chamber technology ensures long-term operational stability even under a full load.
Precise Humidification - BINDER's long-life steam humidification system and the capacitive humidity sensor ensure on-target humidity values.
Other Features Included A Standard
Temperature range - 0 °C to 90 °C
Humidity range - 10 % to 98 % RH
APT.line™- preheating chamber technology
Humidity regulation - with capacitive humidity sensor and vapour humidification
Inner chamber - interior made of stainless steel
Intuitive touchscreen controller - with time-segment and real-time programming
Internal data logger - measured values can be read out in open format via USB
Models In The KMF Series
KMF 115
3.60 cu.ft
MPN
9020-0341
KMF 240
8.72 cu.ft
MPN
9020-0343
KMF 720
24.72 cu.ft
MPN
9020-0345
Additional Options & Accessories (For Full List & Part Numbers See Product Downloads)
BINDER Pure Aqua Service - This water treatment system is used to fully desalinate faucet water for BINDER climate chamber humidification systems.
Depth (plus door handle, I-panel, connection 80 mm) (mm) 730
Wall clearance, rear (mm) 100
Wall clearance, side (mm) 100
Number of doors (ea.) 1
Inner glass door(s) (ea.) 1
Interior dimensions
Width (mm) 600
Height (mm) 483
Depth (mm) 351
Interior volume (l) 102
Racks (number standard/max.) 1 / 5
Load per rack (kg) 30
Permitted total load (kg) 100
Weight (empty) (kg) 127
Temperature data
Temperature range (°C) -10 - 100
Mean warm-up rate acc. to IEC 60068-3-5 (K/min.) 1,3
Mean cooling rate acc. to IEC 60068-3-5 (K/min.) 0,5
Warm-up time from -10 °C to 100 °C (min.) 85
Cooling down time from 100 °C to -10 °C (min.) 240
Max. heat compensation up to 25 °C (W) 150
Climate data
Temperature range (°C) 10 - 90
Temperature variation (±K) 0,3 - 1,0
Temperature fluctuation (±K) 0,1 - 0,2
Humidity range (% RH) 10 - 90
Humidity fluctuation (± RH %) = 2,5
Dew point temperature range (°C) 5 - 80
Max. heat compensation at 25 °C / 90 % RH (W) 30
Recovery time after door was opened for 30 sec
at 25 °C / 60 % RH (min.) 15
at 40 °C / 75 % RH (min.) 8
Electrical data
IP protection class acc. to EN 60529 IP 20
Voltage (± 10 %) 50 / 60 Hz (V) 200 - 240, 1 N ~
Nominal power (kW) 2,0
Energy consumption at 85 °C / 85 % RH (W) 1) 570
Noise level (dB (A)) 52
These values can be used for dimensioning air condition systems.
All technical data are specified for units with standard equipment at an ambient temperature of 25 °C and a line voltage fluctuation of ±10%. The temperature data is determined in accordance to factory standard following DIN 12880, respecting the recommended wall clearances of 10% of the height, width and depth of the inner chamber. Technical data refers to 100% fan speed. All figures are typical average values for series devices. We reserve the right to alter technical specifications at any time.
Speak to us about your requirements and we'll give you a competitive price. Arrange an onsite visit if you're unsure whether a product offers the best solution for your business's needs.